Abstract

Performance of a simple method of particle identification via pulse rise-time discrimination is demonstrated for slow pulses from charge-sensitive preamplifiers with rise times ranging from 10 to 500 ns. The method is based on a comparison of the amplitudes of two pulses, derived from each raw preamplifier pulse with two amplifiers with largely differing shaping times, using a fast peak-sensing ADC. For the injected charges corresponding to energy deposits in silicon detectors of a few tens of MeV, a rise-time resolution of the order of 1 ns can be achieved. The identification method is applicable in particle experiments involving large-area silicon detectors, but is easily adaptable to other detectors with a response corresponding to significantly different pulse rise times for different particle species.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.