Abstract

In order to maintain a constant monochromatic synchrotron X-ray beam height for all energies, the separation between the crystals of a double-crystal monochromator is typically adjusted, via translation of the second crystal, while X-ray energy is varied, via rotation of the crystal pair. The ability to accurately translate the second crystal requires precise knowledge of the separation between the two crystals and, when present, crystal miscuts. Here, a simple method for calibrating the crystal gap from measured variation in the X-ray beam height that eliminates error in the fixed beam offset is provided.

Highlights

Read more

Summary

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.