A reliable, simple, and affordable liquid tensiometer is presented in this paper. The instrument consists of 72 ultrasonic transmitters in a tractor beam configuration that levitates small liquid samples (droplets) in air. Under operation, the instrument imparts a pressure instability that causes the droplet to vibrate while still levitating. Droplet oscillations are then detected by a photodiode, and the signal is recorded by an oscilloscope. The frequency of these oscillations is obtained and then used to obtain the effective surface tension of the sample. The instrument operates at the millisecond scale time (t < 12.5 ms), with very small liquid volumes (∼0.5 μl), and the sample is recoverable after testing. The instrument has been experimentally validated with acetone, ethanol, Fluorinert FC-40, water, and whole milk.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.