Abstract

Measurements of 222Rn (“radon”) in the environment are important in the geosciences and radiation-protection fields. We demonstrate here a simple laboratory-based calibration system to evaluate the efficiency of radon detectors with a reproducibility of about ±2%. The system uses a closed-loop air circulation design with 226Ra adsorbed onto MnO2-impregnated fiber as a radon source. Two RAD7 radon detectors (Durridge Co., Inc.) that were precisely calibrated at Durridge’s in-house calibration facility are used as secondary standards. By parallel analysis of the radon-enriched air within the closed loop, the test RAD7s are assigned a calibration coefficient to be applied to future measurements. We also performed a side-by-side intercomparison with two RAD7s in a high-radon natural environmental setting (limestone cave in Florida) that produced comparable results.