Abstract

This work shows the results obtained by Inmetro’s mass lab implementation of a simple Kibble balance approach for demonstration of its mass measurement principle based on the Planck’s constant. It is based on some modifications of Mimimal Watt Balance work reducing the required electrical measurements by using two loudspeakers mechanically coupled so that one of them acts as a linear actuator in order to perform BL measurements in the moving phase. The mean BL value obtained in weighing phase was 2.66 N/A and in moving phase 2.67 N/A. The maximum error of mass values measured for standard weights in the range from 5 g to 50 g was lower than 5%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.