Abstract
OBJECTIVE: A high-performance liquid chromatography method for the estimation of Linezolid in human plasma was developed and validated. METHODS: Samples (100µµL) were deproteinized with acetonitrile and analyzed using LiChrospher 100, RP18e column with PDA detection at 254 nm. The flow rate of the isocratic mobile phase comprising of 0.1% formic acid in 1000 ml of water and acetonitrile in the ratio of 60:40 (v/v) was set at 1.0 ml/min. RESULTS: The calibration curve ranged from 0.50 to 20.0 µg/ml and was linear. The recovery ranged from 96% to 101%. The accuracy ranged from 98 to 101% and intra- and inter-day relative standard deviation was <4.58%. The method reliably eliminated interfering materials from plasma and R2 was 0.9973. The method described was applied to the determination of plasma LZD concentration in multi-drug-resistant tuberculosis patients who are treated with a dose of 600 mg LZD once daily. CONCLUSIONS: The developed method is suitable for determination of plasma LZD in routine care and considered feasible in less-resourced settings
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.