Abstract

A rapid and sensitive high pressure liquid chromatographic method has been developed for the routine monitoring in serum of the antineoplastic agent fluorouracil, during continuous intraportal administration. Serum spiked with internal standard, bromouracil, was gently shaken for 10 min with ethylacetate at an acidic pH. The evaporated extract was dissolved in the mobile phase containing potassium phosphate buffer. Assays were performed at ambient temperature on a Chrompack LiChrosorb RP 18 column (2 X 100 X 3.0 mm) equipped with a guard column. The retention time of fluorouracil was 2-6 min. The calibration curve was linear from 25 to 2,000 ng/ml. The coefficient of variation was less than or equal to 1.6% for within-run precision and less than or equal to 2.9% for day-to-day precision. The mean recovery of fluorouracil was 57%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.