Abstract
A simple and small-sized direct current glow discharge (dc-GD) ion source has been designed and constructed for direct analysis of solid samples. The glow discharge (about 100 mm3 of internal volume) was coupled to a commercial on-axis time-of-flight mass spectrometer (TOFMS) for fast simultaneous multielement-selective detection. The commercial TOFMS used was originally coupled to an inductively coupled plasma (ICP) ion source. Therefore, the GD source presented in this study was designed aiming at being easily exchanged with the ICP source of a commercial ICP-TOFMS instrument (LECO Corporation). The distance between the cathode (sample) and the sampler cone was kept at about 4 mm.The proposed source is able to analyse planar conducting samples of different sizes with good long-term stability (better than ±5% when isotope ratioing techniques were applied). The spectral mass resolving power was about 1520 (at mass 120 u).Finally, the expected benefits of the fast simultaneous multielemental capabilities of TOFMS detection to carry out direct depth profile analysis of solid samples are demonstrated and discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.