Abstract

A simple flow injection spectrophotometric method for the determination of nitrite is described. Nitrite injected into the flow system reacts with thiourea in acidic medium and the generated thiocyanate ion reacts with Fe(III) in the reagent solution to produce a highly colored product. The influences of chemical and physical parameters including reagent concentrations, sample volume injected, flow rates of the carrier and reagent solutions, reaction coil length and reaction temperature, were studied and optimum values of these parameters were established. Under the optimum conditions, the calibration curve for nitrite was linear over the concentration range 0.36 - 90 microg ml(-1) without preconcentration and over the range 3.8 - 500 ng ml(-1) with a simple online preconcentration step using an anion exchange column. The corresponding detection limits were 0.36 micro ml(-1) and 3.8 ng ml(-1), respectively. Up to 25 samples can be analyzed per hour, with an average relative standard deviation of < or = 1.2%. Interferences by various foreign ions were studied and the method was applied to the determination of nitrite in water and spiked water samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.