Abstract

A simple, fast, accurate and reliable modification of the Schütze-Unterzaucher oxygen determination method is described. Interfering products from the pyrolysis of sulfur and halogen containing compounds are eliminated with a liquid oxygen trap. The carbon monoxide is oxidized with anhydroiodic acid. The products formed in this reaction are absorbed in a CH-type absorption tube and weighed. Experiments with another very fast and efficient method, in which the pyrolysis is carried out in vacuum, are reported.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.