Abstract

A simple, inexpensive mechanical device is described for the direct continuous measurement of mean arterial pressure. It can be assembled in minutes using a few dollars worth of sterile components commonly available in an intensive care unit. Mean arterial pressure determined in patients with this device differs by 1.6 mmHg +/- 1.1 SD from mean arterial pressure simultaneously measured by properly calibrated Hewlett-Packard electronic monitors (N = 5), a non-significant difference. Similar agreement was obtained when the device was tested using a commercial blood pressure system calibrator as the pressure source. This device also allows rapid static calibration of electronic pressure transducers across the full range of human blood pressure using a 3 ml syringe as the pressure source and an aneroid gauge, or mercury column, as the pressure standard.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.