Abstract
This article presents a simple measuring apparatus that is capable of measuring column density and absolute photoabsorption cross sections. The apparatus includes a simple high-temperature heatpipe furnace, an MKS Instruments Baratron, a hydrogen discharge lamp, and an optical detecting system. The heatpipe furnace provides an absorbing medium with a steady total pressure in a closed system so that the column density can be obtained using the ideal gas relation. The apparatus is designed for a working temperature range below 1000 K and a pressure range up to 10 Torr. Encountered experimental difficulties and their found solutions will be discussed in detail. In this work, the absolute photoabsorption cross section of Mg is measured near the 3s threshold using this apparatus. The absolute photoabsorption cross section result at the 3s threshold, which is 2.1±0.3 Mb, is in good agreement with hook method data (2.36 Mb).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.