Abstract

In this work, a simple, selective and rapid flow injection method has been developed for determination of ruthenium. The method is based on its catalytic effect on the oxidation of safranin-O by metaperiodate. The reaction was monitored spectrophotometrically by measuring safranin-O absorbance at λ max = 521. The reagents and manifold variables, which have influences on the sensitivity, were investigated and the optimum conditions were established. The optimized conditions made it possible to determine ruthenium in the ranges of 0.4–20.0 ng/mL (Δ A = 0.2819 C Ru + 1.1840) and 20.0–100.0 ng/mL (Δ A = 0.0984 C Ru + 7.9391) with a detection limit of 0.095 ng/mL and a sample rate of 30 ± 5 samples/h. Relative standard deviation for the five replicate measurements was less than 1.84%. The proposed method has been successfully applied for analysis of ultra trace amounts of ruthenium in real samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call