Abstract

This paper presents a rapid less than 2 min and low-cost method involving the use of alkali solution to capture the acidic gasses from a biogas, thereby providing an estimate of the percentage of non-acidic gasses. Such a method was mentioned in the literature but never fully described or optimized. After sampling an aliquot of gas from bioprocess, gas was injected in a sealed flask with a 3 M NaOH solution, and after equilibrium was obtained, the non-acidic gas volume was measured. The method was first calibrated with certified gasses with an accuracy observed between 98 and 105%. Regarding the validation step, certified standard gas mixtures and nine biogas-laboratory batch reactors were used, the overall accuracy reported was 103 + 3%. This rapid and low-cost method may either be used in laboratory conditions as a quick and low cost alternative to standard analysis equipment or in addition as a routine field control method used on full-scale plants.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.