Abstract

To demonstrate why conventional non-invasive mouthpiece-based fixation has not achieved the expected accuracy and to suggest a solution of the problem. The Vogele Bale Hohner (VBH) head holder is a non-invasive vacuum mouthpiece-based head fixation system. Feasibility and repositioning accuracy were evaluated by portal image analysis in 12 patients with cranial tumors intended for stereotactic procedures, fixated with the newest version (VBH HeadFix-ARC). Portal image analysis (8 patients evaluated in 2-D, 4 patients in 3-D) showed that even in routine external beam radiation therapy, treatment can be applied to within a mean 2-D and 3-D accuracy of under 2 mm (SD 0.92 mm and 1.2 mm, respectively) with cost and repositioning time per patient and patient comfort comparable to that of common thermoplastic masks. These preliminary results show that high repositioning accuracy does not rule out simple and quick application and patient comfort. Paramount, however, is tensionless repositioning via the vacuum mouthpiece.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.