Abstract

A simple and fast liquid chromatography-tandem mass spectrometry method was established and validated for the simultaneous determination of tenofovir alafenamide (TAF) and tenofovir (TNF) in human plasma. A simple protein precipitation procedure was employed to extract analytes from plasma. Chromatographic separation was performed on an Eclipse Plus C18 column utilizing a fast gradient elution starting with 2% of 2 mM ammonium acetate-formic acid (100/0.1, v/v) followed by increasing the percentage of acetonitrile. Detection was performed on a tandem mass spectrometer equipped with an electrospray ionization source operated in the positive ionization mode, using the transitions m/z 477.2 → m/z 346.1 for TAF and m/z 288.1 → m/z 176.1 for TNF. TAF-d5 and TNF-d7 were used as the internal standard of TAF and TNF, respectively. The method was validated in the concentration ranges 1.25-500 ng/mlfor TAF and 0.300-15.0ng/ml for TNF with acceptable accuracy and precision.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.