Abstract

A new method is developed for the determination of energy peak efficiency of 208 keV γ-ray of 237U based on 241Pu/237U secular equilibrium. Plutonium solution was purified to remove Am with Dowex 1 × 2 anion exchange chromatography and the concentration of 241Pu in the purified solution was determined using 242Pu isotope-dilution mass spectrometry on an inductively-coupled plasma mass spectrometry. The solution can be used as calibration source for the determination of energy peak efficiency of 208 keV γ-ray of 237U after 48 days. The method was validated for a planar HPGe detector at the 12 mm above the detector surface. The results showed that this is a simple and economic method for determining the energy peak efficiency of gamma detectors for 208 keV γ-ray of 237U.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.