Abstract

An ultrahigh-vacuum setup for studying the interaction of ions with a solid by the thermodesorption spectroscopy method has been developed. The residual pressure achieved in the setup after heating is <2 × 10−7 Pa (the main component of the residual gas is H2). A sample suspended by two tungsten wires is inserted into the main vacuum chamber through a lock chamber. The sample is heated by thermal radiation from the heater from room temperature to 1600–1700 K. The main features of the setup, the experimental technique, and the calibration of desorption flows are discussed. Examples of results obtained on the setup are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.