Abstract

A setup for measuring the lifetime of free electrons in a neutral buffer gas that contains an electronegative impurity is described. This setup is intended for prompt diagnostics of carcinogenicity of chemical compounds and consists of an ionization chamber with laser photoionization of the cathode, a gas system, and a system for collection and processing of experimental data based on an HBJI45 amplitude-to-digital converter. The techniques for preparing the tested gaseous mixtures and processing the experimental results are described. The results of measurements of the lifetimes and calculated free-electron capture constants for a CO2 + O2 calibrating mixture and mixtures of CO2 with organic molecules are presented. The setup allows measurements of the lifetimes of free electrons in mixtures in the range 0.5–70 μs at measurement errors of 0.2–6%, respectively, and study of two or three potentially carcinogenic substances a week.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.