A new method for biochemical oxygen demand (BOD) determination, which combines the series piezoelectric quartz crystal (SPQC) technique with the growth of a microorganism is presented in this paper. This method needs no immobilization of bacteria and is simple and convenient. When a calibration technique was used for BOD analysis, the detection time was 2.5 h at 37°C. There was a good linear relationship between the frequency shift and BOD value in the range 2.2–11 mg/L and the regression equation was ΔF = 64.10 + 11.23[BOD]. The proposed method was compared with the conventional BOD5 method. This method was more rapid than the BOD5 method and the results obtained by the former were in good agreement with those obtained by the latter. The experimental conditions are also discussed in detail.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.