Abstract

The present study reports on a liquid chromatography-tandem mass spectroscopy method for the analysis of Cilostazol drug substance in its formulation. The materials utilize a Zorbax SB C18 analytical column with electrospray ionization and a triple quadrupole mass detector in multiple reaction monitoring (MRM) mode, resulting in a short analysis of 1.5 min runtime and the analyte elute at 0.76 min. Change from SIM to MRM mode resulted in increased sensitivity and low quantitation limit which in turn produces a detection limit of 1 ng/ml. The obtained method was validated in a range of 5-100 ng/ml with a correlation coefficient of 0.9991. The percentage recovery of the present method was found to be in the range of 94% to 100%. The developed method was validated as per the ICH guidelines.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.