Abstract

Pentosidine is an advanced glycation end-product (AGE) appearing when arginine and lysine residues in proteins are cross-linked with carbonyl derivatives. This paper presents an improved method for the synthesis of pentosidine and reversed-phase chromatography of this substance with fluorometric detection that enables sensitive (0.01 pmol/mg protein) and specific determination of pentosidine in plasma. Separation is done twice on the same C 18 Vydac 218TP54 column, first with trifluoroacetic acid and next with heptafluorobutyric acid as ion pair. The inter-day coefficient of variation is 6.4% at pentosidine concentration in plasma of 25 pmol/mg protein and 8% at 1.7 pmol/mg protein. Spectral properties of pentosidine exploited during identification of the substance with UV absorption and fluorescence detectors are described. Maximum of absorbance was observed at 325 nm, maximum fluorescence at λ ex/ λ em=330/373 nm. The method may prove useful for the study of processes associated with generation and accumulation of pentosidine in the body as a marker of AGE production in healthy subjects and patients with chronic renal failure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.