Abstract

The experimental work involved the efficiency determination of three coaxial hyperpure Ge detectors, using point-like sources placed at various distances from the top cover of the detectors. Curves were fitted to the experimental data and a microcomputer routine was developed to calculate the efficiencies of cylindrical sources, including self-attenuation correction. Tested at various source and distance conditions, the procedure showed results sufficiently consistent for routine k0-INAA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.