Abstract
A semi‐automated, 96‐well solid phase extraction (SPE) method combined with high performance liquid chromatography (HPLC) was developed, validated, and implemented to demonstrate its high sample throughput utility for the determination of a selective gamma amino‐butyric acid (GABA)‐A receptor agonist in human and rat plasma. A Tomtec Quadra 96 system was utilized to perform the typical 96‐well SPE steps, such as plate conditioning and washing, sample loading, drug and internal standard elution, dilution, and liquid transfers. The extracted and diluted eluents were analyzed by HPLC with fluorescence detection with excitation and emission wavelengths at 256 and 447 nm, respectively. Precision and accuracy for standard curve and quality control samples in human and rat plasma were all within acceptable limits. Post‐dose plasma samples from three rats were prepared and analyzed using this semi‐automated 96‐well SPE method. Full details of the analytical methodology, method validation, and the results of analyses of rat plasma samples are presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Liquid Chromatography & Related Technologies
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.