Abstract
A self-reciprocity method is described for calibrating a spherical transducer in a spherical shell. The reciprocity constant is calculated using Green's function and the electroacoustic reciprocal principle in the spherical shell. A sensitivity correction is developed by calculating the transducer's receiving force in different fields. An experimental measurement setup is described to calibrate transducers with diameters of 20 and 30 mm in the frequency range from 25 to 63 kHz, calibrated in a 500 mm diameter spherical shell with a thickness of 2 mm. The largest discrepancy in calibration results between the self-reciprocity and three-transducer spherical-wave reciprocity method is around -1.4 dB. The self-reciprocity calibration method, with a smaller volume required than the three-transducer spherical-wave reciprocity method, has a potential advantage in calibrating the transducer at high hydrostatic pressure and varying water temperatures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.