Abstract
National Institute of Metrology (NIM) has set up a new large-area tungsten carbide–carbon (WC-C) high-temperature fixed-point (HTFP) blackbody system as the direct radiation source for measuring spectral irradiance. The point of inflection (POI) of the HTFP melting plateau curve was important to determine the liquidus point and applied as the reference point in temperature calibration. This paper presented a selective multiple fit method. The maximum discrepancy of different factors for the new method and traditional methods were 0.001 K and 0.633 K which led to 0.0003% and 0.20% spectral irradiance errors at 500 nm. In addition, when used to calculate POIs of WC-C and Re-C small-area HTFPs, the maximum deviation between the new method and the average value of traditional methods were −0.004 K and −0.03 K. This method was verified to be insensitive to the influencing factors and more effective for large-area HTFPs applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.