Abstract

The Palmes’ tube, the first diffusive sampler incorporating a fixed path length, has received wide usage for the sampling of a large number of gaseous pollutants. But despite numerous previous studies, questions remain regarding the accuracy of these inexpensive, simple-to-construct, open-ended samplers. Here the mass transfer resistance in a Palmes’ diffusive sampler was measured using the loss of cyclohexane from a Palmes’ tube containing liquid cyclohexane at its base. The average loss rates, at factorial combinations of five air incidence angles evenly spaced from 270° to 90°, and five air speeds from 0.5 m/sec to 2.5 m/sec ranged from 46% to 121% higher than rates calculated from the physical dimensions of the sampler, proving the need to calibrate these samplers rather than relying on a theoretical calculation. The mass transfer resistance was nearly constant when the airflow was perpendicular to the sampler and sufficiently high to avoid stagnation, a finding that may explain the widespread acceptance of the results obtained using this sampler.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.