Abstract

A sample cell and low-temperature cooling unit design suitable for x-ray absorption spectroscopy (XAS) of oxygen and/or water sensitive compounds is presented. The cell is a simplified and optimized modification of a previously suggested design. The low-temperature accessory is specifically designed for good thermal isolation and equilibration at liquid nitrogen temperatures, eliminating the need for a more expensive commercial cryostat. The entire construction can be directly attached to an ion chamber and either a standard Lytle detector or energy-discriminating germanium-element solid-state detector via helium gas-purged adaptor flanges. The apparatus was tested at liquid nitrogen temperatures and gave good temperature stability within ±0.1 K. The XAS signal from an air-sensitive potassium/pyrene complex at 100 K taken at the X-6B beam line at the National Synchrotron Light Source facility at Brookhaven National Laboratory is presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.