Abstract

At the lowest temperatures achieved in dilution refrigerators, ruthenium oxide resistance thermometers often saturate and therefore lose their sensitivity. In an effort to extend the range of such temperature sensors, we built a thermometer which maintains sensitivity to 5 mK. A key component of this thermometer is an in situ radio frequency filter which is based on a modern rf absorption material. We show that the use of such a filter is only effective when it is encased in the same rf-tight enclosure as the ruthenium oxide sensor. Our design delivers an attenuation level that is necessary to mitigate the effects of parasitic heating of a fraction of pW present in our circuit. Furthermore, we show that the likely origin of this parasitic heating is the black body radiation present within the experimental space of the refrigerator. We found that the equilibration time of the thermometer increases rapidly as the temperature is lowered; below 5 mK this thermometer becomes impractical because of the prohibitively long equilibration times.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.