Abstract

Quality control principles, initially designed for manufacturing, have been applied to clinical chemistry and later to serological testing for infectious diseases. Methods for setting control limits have been included using the effects of assay performance on clinical outcomes. However, this approach assumes that the reactivity of patient samples change proportionally with the results of quality control samples tested in the same assay. Although this may be the case for clinical chemistry, this assumption has never been tested for serological assays. During a 9-month period, 177910 and 185684 donor test results for HBsAg and anti-HIV, respectively, were analysed and compared with 712 HBsAg and 710 anti-HIV results obtained from a single batch of a multimarker quality control sample obtained from the same assays. Results of the analysis showed that the negative donor results did not change in proportion with the changes experienced by the quality control sample results as determined by regression coefficients, particularly when results obtained from different reagent batches were compared. There were insufficient results obtained from positive samples for a similar analysis. However, investigations show that, even if the reactivity of positive donor samples changed in proportion with the change in reactivity of the quality control sample, few false-negative donor test results would have occurred.