Abstract

The Trifecta valve has externally mounted leaflets; it differs from classic internally mounted valves (e.g., Carpentier-Edwards [C-E]). We evaluated post-implantation hemodynamics and clinical outcomes of these bioprostheses in small aortic annuli. From January 2015 to April 2019, 490 patients who underwent aortic valve replacement (AVR) were reviewed retrospectively. Altogether, 183 patients received 19 or 21 mm diameter C-E (n = 121) or Trifecta (n = 62) prostheses. To minimize confounding variables, we performed propensity-score matching analysis. The mean transvalvular pressure gradient (TVPG) was significantly lower in the Trifecta than in the C-E group at discharge (12.9 ± 4.8 vs. 15.0 ± 5.3 mmHg, p = 0.044). TVPG change over time was not significantly different between groups (p = 0.357). Left ventricular mass index decreased postoperatively (reduction: C-E, 28.1%; Trifecta, 30.1%, p = 0.879). No late mortality, severe patient–prosthesis mismatch, moderate-to-severe paravalvular leakage, structural valve degeneration, or valve thromboses were observed. Freedom from valve-related events at 3 years were similar for C-E (97.9% ± 2.1%) and Trifecta (97.7% ± 2.2%) patients (log-rank p = 0.993). Bioprosthesis design for small annuli significantly affected TVPG immediately after AVR. However, hemodynamics over time and clinical outcomes did not differ between the two designs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.