Abstract

The authors investigated the causes for surgical reexploration in patients with complex regional pain syndrome Type II who received initial relief of pain from implantation of a peripheral nerve stimulator (PNS). The authors reviewed the charts of 11 consecutive patients who underwent a total of 27 PNS-related operations at one institution. Duration of follow up ranged from 5 days to more than 24 months. Of 11 patients who received PNS implants, seven (64%) required one or more additional surgeries to relocate the PNS because initial pain relief following stimulation was lost and not restored by changing pulse generator settings. Loss of analgesia was attributed to migration of the sutured electrode strip paddle (nine [33%] of 27 surgeries), infection (four [15%] of 27), and the need for placement in an alternative location (three [11%] of 27). Although infection is attributable to surgical technique, most complications requiring repeated surgery (nine [33%] of 27) are caused by equipment design. Changes in PNS design or in implantation technique might substantially reduce the need for reoperation after PNS implantation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.