Abstract
AbstractA spectroelectrochemical detector for flow injection and liquid chromatography is described. The detector design has dimensions of a standard 1 cm sample cuvette with a replaceable reticulated vitreous carbon working electrode. No special cell holder or modifications to the spectrophotometer are required to use the detector. Chronocoulometry is used to indicate a trend from thin‐layer behavior toward semi‐infinite diffusion behavior as the electrode pore size increases. Ferro‐/ferricyanide solutions were used to evaluate the flow injection characteristics of the detector. A sampling rate of 180 injections per hour was used. Linear calibration curves in the micromolar range were achieved with a reproducibility of 2–7%. A mixture of phenol, chlorophenols, and nitrophenols was used to investigated the simultaneous electrochemical and optical response for liquid chromatography.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.