Abstract

AbstractA spectroelectrochemical detector for flow injection and liquid chromatography is described. The detector design has dimensions of a standard 1 cm sample cuvette with a replaceable reticulated vitreous carbon working electrode. No special cell holder or modifications to the spectrophotometer are required to use the detector. Chronocoulometry is used to indicate a trend from thin‐layer behavior toward semi‐infinite diffusion behavior as the electrode pore size increases. Ferro‐/ferricyanide solutions were used to evaluate the flow injection characteristics of the detector. A sampling rate of 180 injections per hour was used. Linear calibration curves in the micromolar range were achieved with a reproducibility of 2–7%. A mixture of phenol, chlorophenols, and nitrophenols was used to investigated the simultaneous electrochemical and optical response for liquid chromatography.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.