Abstract

An NMR Q-switch was designed and constructed specifically for use with low-field NMR apparatus. This featured a comparatively simple resistive damping design. It served to reduce the r.f. probe ring-down time, and hence reduced the signal acquisition delay from 25 ms to 9 ms, on an Earth’s magnetic field NMR system. The advantage of this earlier acquisition was demonstrated for both an aqueous suspension of iron oxide particles and using an NMR flow meter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.