Abstract

The LPM (Van Straten Medical, Netherlands) proximal interphalangeal joint replacement was introduced to the United Kingdom in October 2000, and 257 implants were sold. We have audited the results of this prosthesis, and obtained data for 164 of the 257 implants. Forty-seven (29%) of the 164 have been revised and another 33 (20%) are showing clinical and radiological signs of failure at a maximum follow-up of 6 years. Massive osteolysis leading to aseptic loosening is the commonest cause of failure. We believe that all patients who have had this prosthesis inserted should be kept under regular clinical and radiographic review.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.