Abstract

Luminex single antigen bead (SAB) assays have become an essential tool in monitoring the status of antibody to the Human Leucocyte Antigen (HLA) of patients both before and after transplantation. In addition SAB data is used to aid risk stratification to assess immunological risk of humoral rejection in solid organ transplantation (CTAG/BTAG guidelines) [1]. Increasingly laboratories are reporting false negative results at high antibody titre due to a prozone effect.Here we report a case study where the prozone effect led to a false negative antibody result that could have resulted in adverse outcome. We describe a method to reliably remove the prozone effect through heat inactivation and the addition of Ethylenediaminetetraacetic acid (EDTA) to the Luminex wash buffer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.