Abstract
In the present paper, the assessment of vortex-assisted dispersive magnetic solid-phase extraction using amino-functionalized mesoporous combined with direct magnetic sorbent sampling (DMSS) in flame or furnace atomic absorption spectrometry (FAAS or FF-AAS) was demonstrated for highly sensitive silver determination in water samples. The developed method showed significant enrichment factors compared to conventional pneumatic nebulization by FAAS, 607 for DMSS-FF-AAS and 114 for DMSS-FAAS. The analytical curve showed linearity in the range from 5.0 to 70.0µg L- 1 and 1.0 to 15.0µg L- 1 and limits of detection of 0.59 and 0.09µg L- 1 for DMSS-FAAS and DMSS-FF-AAS, respectively. The intra and inter-day precision evaluated as a percentage of the relative standard deviation (RSD,%) ranged from 1.89 to 4.71% for levels of 25.0 and 65.0µg L- 1. The method was applied in different kinds of water samples without matrix effects, yielding recovery values from 90 to 110%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Bulletin of environmental contamination and toxicology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.