Abstract

Serial measurements of absolute CD4+ T-lymphocyte counts are required to initiate and gauge response to therapy and monitor disease progression. Hence, there is an urgent need to evaluate the accuracy and validity of low-cost CD4+ T-cell count assays. Tripotassium EDTA blood specimens from HIV-infected individuals were studied using a novel flow cytometric assay (EasyCD4 assay; Guava Technologies, Hayward, CA) in comparison with standard flow cytometry (FACSCount; Becton Dickinson Immunocytometry Systems, San Jose, CA). The sensitivity, specificity value by EasyCD4 assay in enumerating absolute CD4+ T-cell counts of less than 200 cells/microL were 95% and 100%, respectively. Bland-Altman analysis showed close agreement, with the EasyCD4 assay yielding CD4+ T-cell counts a mean difference of -26 cells/microL (95% confidence interval, -96 to 44 cells/microL) higher than by flow cytometry. Our data suggest that EasyCD4 assay could be a useful alternative assay to conventional flow cytometry, may be appropriate for use in resource-limited settings.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.