Abstract

PurposeTo evaluate the current beliefs about the ways to reduce target registration error (TRE) values in image guided Sinus surgery by rearranging the fiducial configuration, and investigating the best configurations for various surgical fields in a phantom study. MethodsA new CT-compatible skull phantom consisting of implanted targets was designed to enable direct measurement of TRE in four fields of sinus surgery, Frontal, Ethmoid, Sphenoid and Maxillary. The effects of different landmark configurations on TRE values, measured by the Parsiss-IV navigation system were investigated to find the best landmark arrangement for each region, and compared to the TRE prediction formula to assess the clinically accepted landmark selection approaches based on this formula. ResultsIt was shown that smaller values of TRE could be attained by arranging the center of the fiducials to be more focused on the surgery target. The addition of more fiducials and keeping non-linear arrangement of landmark would not necessarily decrease the TRE value. ConclusionOptimizing the landmark configuration is important for increasing the localization accuracy in image guided sinus surgery. The common beliefs accepted in the clinical community about the ways to reduce the TRE are very general and should be adapted to specific field of image guided surgery.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.