Abstract

A reference tristimulus colorimeter has been developed at NIST with a transmission-type silicon trap detector (1) and four temperature-controlled filter packages to realize the Commission Internationale de l'Eclairage (CIE) x(λ), y(λ) and z(λ) color matching functions (2). Instead of lamp standards, high accuracy detector standards are used for the colorimeter calibration. A detector-based calibration procedure is being suggested for tristimulus colorimeters wehre the absolute spectral responsivity of the tristimulus channels is determined. Then, color (spectral) correct and peak (amplitude) normalization are applied to minimize uncertainties caused by the imperfect realizations of the CIE functions. As a result of the corrections, the chromaticity coordinates of stable light sources with different spectral power distributions can be measured with uncertainties less than 0.0005 (k=1).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.