Abstract

The calibration of low current meters in the range of picoamperes up to microamperes is important for the electronics industry, for the photometry area, and especially for the ionizing radiation area. In Brazil, traceability is limited to ten microamperes. In this paper, a reference system using Ohm’s law method is presented. It was possible to calibrate an electrometer between 2 picoamperes and 20 microamperes, with uncertainties between 0.70 % and 0.000060 %, respectively, as well as to validate the developed system and Ohm’s law method by comparing it with the method used to calibrate values above 10 microamperes, in an intralaboratory comparison.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.