Abstract

One major factor of red blood cell damage induced by artificial heart valves is the magnitude of turbulent shear stresses in the flow field. An often-cited threshold for hemolysis is 400 N/m 2 (Sallam and Hwang, Biorheology 21 (1984) 783). This value, however, was measured with a one-component laser Doppler anemometer and was not calculated from the major principal Reynolds shear stress, therefore underestimatng the threshold. This study performed flow field measurements under similar conditions, using a two-component laser Doppler anemometer and determining major principal Reynolds shear stress, to reevaluate the hemolytic threshold as 800 N/m 2 with an exposure time of 1 ms. The Kolmogorov length scales, approximately 9 μm, are capable of inflicting direct damage to the red blood cells. These results will serve as a more accurate reference in future heart valve design and testing.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.