Abstract

Aiming at the problem of timeliness of CBCT reconstruction, a CBCT fast short scan reconstruction method is proposed. At the same time, the image reconstruction process in which a new attenuation compensation algorithm is applied to improve image quality. When performing FDK three-dimensional reconstruction of a single-frame acquisition image, the Parker-weighted image is calculated in real time, and a new attenuation compensation algorithm is applied in the back projection process to complete the short scan Parker-weighted reconstruction. This method simulates the CBCT synchronous acquisition and reconstruction process by establishing collection and reconstruction threads. Under the premise of satisfying the reconstruction quality, the reconstruction can be completed within 1 to 2 seconds after the patient collection is completed, which achieves the purpose of real-time.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.