Abstract

This work is focused on hygrothermal control problem of closed newborn incubators. Such incubator promotes a controlled micro-climate, with small heat transfer between the premature and the environment, which promotes a healthful environment. In this context, the main hygrothermal informations are temperature, relative humidity and partial vapor pressure where the last one is closed related with the water loss by transpiration. A model predictive control scheme for this problem is presented and real time closed-loop control examples validate the proposed method in such a context. The identification procedure, a need in model based control schemes, is based on the use Laguerre functions and the experiments are performed on a laboratory pilot plant (full scale) which was built to evaluate control algorithms in incubators.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.