Abstract

B-mode ultrasound imaging guidance of cannulas can be compromised by noise, artifacts, and echogenicity that is not distinctive from that of surrounding anatomy. We have modified a venovenous extracorporeal membrane oxygenation cannula by embedding piezoelectric crystals into each of its 3 blood flow ports. Each vibrating crystal acoustically interacts with a Doppler imaging signal and produces an instantaneous color marker. The aim of this study was to compare identification of the extracorporeal membrane oxygenation cannula ports by B-mode imaging versus the color Doppler marker. Unlike B-mode imaging, the color Doppler marker identified the corresponding port even in highly challenging closed-chest scans in anesthetized pigs. The method could improve guidance accuracy of cannulas by ultrasound scans.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.