Abstract

To highlight the importance of preventing visceral injury due to prominent anterior implants at the spinal column. A 52-year-old man with cord compression was treated with a T9/10 discectomy and instrumented fusion via a right thoracotomy and trans-pleural approach. Postoperatively, he had improved lower limb numbness. However, after a bout of coughing, there was sudden increase in chest drain output, hemodynamic instability and massive fluid collection in the right chest cavity. Emergency re-exploration of the thoracotomy was performed and a 6cm laceration of the right postero-medial diaphragm was identified as the bleeding source and was found to be in close proximity with the locking nut of the anterior implants. The laceration was repaired and a soft synthetic patch was used to cover the implants. Postoperatively, the hemothorax resolved and the patient recovered from the neurological deficit. Prevention of diaphragmatic injury can be performed using lower profile and less sharp-edged implants. Implant coverage with a soft synthetic material is necessary if unable to perform direct repair of the parietal pleura over the implants.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.