Single cell sorting is commonly used for ensuring monoclonality and producing homogenous target cell populations. Current single cell verification methods involve manually confirming the existence of single cells or colonies in a well using a standard light microscope. However, the manual verification method is time-consuming and highly tedious, which prompts a need for an accurate and rapid detection method for verifying single cell sorting capability. Here, we demonstrate a rapid single cell sorting verification method using the Celigo Image Cytometer. Calcein AM-stained Jurkat cells and fluorescent beads are sorted into 96-well half area microplates using the MoFlo Astrios EQ. Whole well bright field and fluorescent images are acquired and analyzed using the image cytometer in less than 8 min. The proposed single cell verification detection method in multi-well microplates can allow for quick optimization of FACS instruments at flow core laboratories, as well as improvement of downstream biological assays by accurately confirming the presence of single cells in each well.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.