Abstract

A rapid, semi-automated method for determining hexoses which are bound to components of whole blood; e.g., hemoglobin, is described. The method is an improvement on similar lengthy manual methods (1,2) in which 5-hydroxymethylfurfural is produced and measured. Heating of diluted blood samples at 15 lb./sq.in. above atmospheric pressure in a pressure cooker for 20 minutes is shown to give the same results as those obtained with other methods in which 2-5 hours heating at atmospheric pressure is specified (1,2,3). The trichloracetic acid precipitation step needed in the manual methods is avoided by continuous flow dialysis on a Technicom Auto Analyzer II system. Various experimental factors have been studied and comparisons made with a commonly-used ion-exchange resin column technique. Two different methods of calibration, one using pure 5-hydroxymethylfurfural standards, and the other using patients' samples with pre-determined, known values as standards, are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.