Abstract

Published gas-liquid chromatographic (GLC) methods for the determination of nicotine and cotinine have proved impractical for the analysis of a large number of clinical samples. Significant improvements over other methods have been achieved, being low sample volume (0.5 mL plasma), rapid two-step extraction from plasma, no evaporation step, and good separation. The lower limits of sensitivity for nicotine and cotinine were 1 and 5 ng/mL, respectively. The method was validated by the analysis of plasma samples from cigarette-smoking volunteers. The method described permits the quick, routine determination of nicotine and cotinine in a large number of samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.