Abstract

A procedure for determining total uranium in urine using a pulsed-laser Kinetic Phosphorescence Analyzer (KPA) is described. The chemical procedure is simple and rapid, and the operation of the KPA is straightforward. An inter-laboratory comparison (measuring total uranium by using fused pellet fluorometry, alpha spectrometry, and the KPA) demonstrates accuracy. The detection limit for this method, of 0.06 ng/mL, is readily achieved. The relative standard deviation (RSD) of the measurements at the 1.0 ng/mL level is 5%. Interference studies, which include selected elements found in urine, are discussed. Potential interferences in other matrices such as soil, milk, and water are also examined.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.